Features:
- The instrument measurement takes 30-60 seconds using a solvent for the extraction and NDIR method.
- The use of a quartz sample cell keeps the instrument clean from possible spills.
- Oil is extracted using the extremely stable S-316 solvent extractant.
- Users can display analysis results in three different units-mg/l, mg/kg, and Abs
- The instrument has an easy-to-read backlit liquid crystal display.
- RS-232C interface is standard equipment.
- Solvent extraction, NDIR
- Analysis (Infrared spectroscopy)
- Measurement range: Water:0 to 200 mg/L, Soil:0 to 1000 mg/kg, Parts:0 to 1 Abs.
- Measured value; 3 1/2digits LCD with back-light
- Message; Character display LCD with back-light (16 x 2 characters)
- Function: Self diagnostics, Auto hold function, Calendar clock
- Ouput: RS-232C and centronics printer port
- Ambient Temp: 0 to 40℃
- Power: 100 to 240 V AC ±10%,50/60 Hz, 60 VA
- Dimensions: 200 (H) x 250 (W) x 285 (D) mm
- Mass: Approx. 5 kg
The OCMA-310 is an automated portable unit for monitoring oil content in both fresh water and saltwater. The one-touch monitoring process is fully automated, from sample extraction to analysis and sample expulsion. The OCMA-310 utilizes an S-316 extraction solvent to ensure superior oil particle stability and infrared absorptiometry for sample analysis. Using 0mg/l -200mg/l dynamic ranging, the unit is capable of analyzing both high and low concentrations. The OCMA-310 provides swift, highly accurate analysis of the oil content of both fresh water and saltwater as well as that of such fluids as those found at factories and wastewater treatment plants.
Download
Features:
- A microcomputer provides automated analysis of the samples at the touch of a button.
- Dynamic ranging provides analysis from 0mg/l-200mg/l without manual switching.
- The use of an S-316 extraction solvent ensures the highest levels of oil particle stability.
- A digital display eliminates the misreading of analysis results. A visual display alerts users to the unit's operating status.
- The use of a proprietary syringe ensures a fixed amount of sample water and solvent.
- Measurement principle: Solvent extraction, NDIR analysis (Infrared spectroscopy)
- Measurement range: 0mg/l ~ 200mg/l dynamic ranging
- Measurement: Manually controlled
- Calibration: One touch calibration after the calibration standard is introduced to the instrument
- Extraction Solvent: S-316 solvent
- Sample/Solvent Volume: 20 mL, ratio of sample to solvent is 2:1.
- Extraction Method: Built-in extractor for oily water samples
- Display
Measured value; 3 digits LCD with backlight
Message; Character display LCD with back-light (16×2 characters)
- Ambient Temperature: 0 to 40 ℃
- Power: 200 to 240 V AC ±10%,50/60 Hz, 120 VA
- Dimensions: 200(W)×315(D)×342(H) mm
- Mass: Approx. 7 kg